Enhancement of Image Quality Using Detector Filter
Lim Jong-Nam

Kim Hyung-Tae
Kim Min-Hye
Chon Kwon-Su
Abstract
Radiation dose to patient is unavoidable when diagnosis is carried out using X-ray. Radiation diagnosis using dual energy X-ray was examined to verify the possibility of medical applications by SNR and image scoring. The dual energy X-ray was realized by combining together two image plates and filter of 0.5 mm thick Cu or Al. Under one X-ray exposure, contrast enhanced image was obtained using two images of image plates. The enhanced image showed higher SNR and image score compared to the first image which was the image recorded with the first image plate. The dual energy X-ray technique would be a very useful method for obtaining higher SNR image and for realizing very low dose, and could be applied to medical applications.
KEYWORD
Radiography, Dual Energy X-ray, Radiation Dose, SNR