KMID : 1151820160100080629
Journal of the Korean Society of Radiology
2016 Volume.10 No. 8 p.629 ~ p.635
Analysis of the Dental Implants MRI Artifacts by Using the ACR Phantom
Shin Woon-Jae

Abstract
ACR phantom for quality control of magnetic resonance imaging equipment can evaluate magnetic resonance imaging picture quality through various structures within the phantom. In this study, percent Signal Ghosting and Slice position accuracy of imaging could be analyzed by attaching implant and the wire for correction of tooth using ACR phantom in Head coil of 3.0T equipment. In the T1 weighted imaging of the first slice and the eleventh slice of implant, the slice position accuracy appeared to be good in ingress bandwidth 300, and it was good in ingress bandwidth 130 when wire for correction was attached. Percent Signal Ghosting in the seventh slice of SE T1 weighted imaging, implant and wire for correction added all appeared to be good when ingress bandwidth was 230. It is thought that in case of implant dental prosthesis patients in brain exam using magnetic resonance imaging, optimum image can be obtained by changing ingress bandwidth.
KEYWORD
Magnetic Resonance Imaging, ACR Phantom, Implant, Artifacts