Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1151820170110020087
Journal of the Korean Society of Radiology
2017 Volume.11 No. 2 p.87 ~ p.94
A Monte Carlo Study of Dose Enhancement with kilovoltage and megavoltage photons
Hwang Chul-Hwan

Im In-Chul
Kim Jung-Hoon
Abstract
Monte Carlo simulations were used to assess dose enhancement effects for 60-, 90-, 120-, and 150-kV X-rays, and for 6- and 15-MV X-rays. The MCNPX code was used for a computer simulation of the ICRU slab phantom, and gold, gadolinium, and iron oxide (Fe2O3) were employed as dose enhancement agents. In consideration of the buildup region of the incident energy, agent concentrations of 5, 10, 15, and 20 mg/g were inserted on the surface of the phantom at a depth of 5 cm. Based on baseline values obtained in the absence of dose enhancement agents, a quantitative analysis was performed by evaluating depth-dependent changes in the absorbed energy and the dose enhancement factor (DEF). A higher concentration of dose enhancement agents led to a greater dose enhancement effect with iron oxide, gadolinium, and gold in descending order. For kilovoltage (kV) X-rays, as the incident energy was decreased and as the energy became closer to the ionization potential of the atoms in the enhancement agent, the dose enhancement effect increased. In the megavoltage (MV) X-ray range, dose enhancement was higher at 6 MV compared with 15 MV. However, the overall dose enhancements were significantly lower compared to the results obtained with kV X-rays.
KEYWORD
Monte Carlo, Dose enhancement, therapeutic ratio
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)