Noise Changes in Image Plate
Choi Seok-Yoon

Im In-Chul
Abstract
It is possible to do not use the Image Plate for a long time in the clinic. for the acquisition of reliable images the decommissioning process is very important. In this study, we would like to look at complex noises that may occur during Image Plate use depending on the frequency and time of day. first, we got the image from computer radiography. second, We calculate the noise changes using Peak Signal to Noise Ratio(PSNR) and Noise Power Spectrum(NPS). finally, we suggest that remove the noise from the IP for more than 4 hours. and It turned out to be better to remove the noise more than once. this study will be a quantitative reference for Image Plate management and use in the clinical trial.