Journal of the Korean Society of Radiology
2018 Volume.12 No. 4 p.497 ~ p.503
A Study on Improvement of Image Quality Decrease due to Tooth Restoration in Facial CT
Kim Hyeon-Ju

Yoon Joon
Abstract
The purpose of this study was to investigate the degree of image degradation and the improvement of image quality caused by the density difference between the orthodontic filling material and the surrounding anatomical structure during the examination of the facial CT by quantitative and qualitative analysis. The teeth were scanned using 64-MDCT (Discovery 750 HD, GE HEALTH CARE, Milwaukee, USA). The teeth were scanned and compared according to tube voltage, silicone application, and MAR application. As a result, 10.36% CT value decreased at 140 kVp and 5.81% decrease at the application of silicon material. As a result of the qualitative evaluation, it was evaluated that 7 of the 10 observers and 3 of the acceptors were applied to the MAR algorithm. Therefore, it is possible to reduce the unnecessary burden on the radiation exposure dose as well as to reduce the loss of image data by reducing the high density artifacts, as well as the inspection parameters used in the current clinical application and various algorithms that can reduce the high density artifacts. It can be expected to provide a lot of image information.
KEYWORD
CT value, dental restoration, silicone, kVp, MAR