Journal of the Korean Society of Radiology
2018 Volume.12 No. 5 p.645 ~ p.650
Evaluation of Area Dose Product and Image Density according to the Variable Tube Current
Yun Young-Woo

Je Jae-Yong
Abstract
This research aims at measuring images density of according to DAP(dose area product), and suggesting the need to quality control of exposure dose. When tube voltage was fixed as 80 kVp and tube current was set as 1, 25, 50, 80, and 100 mAs, with the increase of DAP from 25 mAs to 50 mAs, the dose also rose 1.88 times as much as before, and with the increase from 50 mAs to 100 mAs, it got 2.05 time higher than before. However, the images density obtained as film grew as much as 48% with the increase from 25 mAs to 50 mAs, and 29% with the increase from 50 mAs to 100 mAs. In addition, it has been found out that the higher the DR images density got from 25 mAs to 50 mAs, the bigger it became by 12%, and that it got bigger by 30% with the increase from 50 mAs to 100 mAs. In other words, the differences in the image density by the increase of the dose with the digital imaging equipment in a proper condition was proved to be less than in the film images. Based on the results of this research, medical institutions using a digital imaging equipment are expected to be able to reduce exposure dose of each region of interest than now through the quality control of radiation dose.
KEYWORD
tube current, film, digital image, density, dose area product