KMID : 1151820190130040605
|
|
Journal of the Korean Society of Radiology 2019 Volume.13 No. 4 p.605 ~ p.611
|
|
Comparison of X-ray Image Quality Between Multi-Function Device(MFD) and Weight Bearing Platforms(WBPs)
|
|
Gil Jong-Won
Lee Kwang-Sung
|
|
Abstract
|
|
|
The purpose of this study is to manufacture a multi-function device (MFD) which can be applied to various types of weight-bearing view of the lower leg, and to compare the results with the images from the existing weight-bearing platforms (WBPs), thereby suggesting a clinical utilization. The MFD was manufactured, by considering the minimum adjustable heights of the platform for weight-bearing foot/ankle, platform for hindfoot alignment view, and X-ray tube of the X-ray device. A foot/ankle phantom was used to take the images of weight-bearing lateral foot in MFD and WBPs to compare the resolutions of the X-ray images using a quick modulation transfer function (MTF) program. Between both the images taken from the MFD and WBPs, there was no statistically significant difference found in the mean cycles per pixel (C/P) and the lines per image height (LPH) of the 50%-Contrast Spatial Frequency (MTF50), and 10-90% of Maximum Energy Rise Distance (10-90%), where p>0.05. The MFD is suggested for its clinical trial as a useful positioning device that can secure the patient's safety and manifoldly perform various inspections. Also, the recommendation of the positioning device as a policy can activate dedicated manufacturers, while also improving the quality of medical services.
|
|
KEYWORD
|
|
Hindfoot Alignment View, Multi-Function Device, Positioning Device, MTF, Weight Bearing Platform
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|