Evaluation of Dose Distribution of 6 MV X-ray using Optical Dosimetry
|
|
Kim Sung-Hwan
|
|
Abstract
|
|
|
In this paper, we developed optical dosimetry system with a plastic scintillator, a commercial 50 mm, f1.8 lens, and a commercial high-sensitivity CMOS (complementary metal-oxide semiconductor) camera. And, the correction processors of vignetting, geometrical distortion and scaling were established. Using the developed system, we can measured a percent depth dose, a beam profile and a dose linearity for 6 MV medical LINAC (Linear Accelerator). As results, the optically measured percent depth dose was well matched with the measured percent depth dose by ion-chamber within 2% tolerance. And the determined flatness was 2.8%. We concluded that the optical dosimetry system was sufficient for application of absorbed dose monitoring during radiation therapy.
|
|
KEYWORD
|
|
Optical Dosimetry, Plastic Scintillator, CMOS Camera, Integral Dose, Percent Depth Dose
|
|