KMID : 1151820200140050695
|
|
Journal of the Korean Society of Radiology 2020 Volume.14 No. 5 p.695 ~ p.702
|
|
Sterilization Effect by the Nd:YAG Laser pulse
|
|
Jeong Kyeong-Hwan
Jung Dong-Kyung Park Jeong-Kyu Choi Hyun-Woo Seo Jeong-Min
|
|
Abstract
|
|
|
The use of medical devices is increasing due to the development of medical technology. Among medical devices, it is often used in the human body for graft and treatment. Therefore, in medical institutions, various sterilization methods according to the type and material of medical devices are applied to prevent infection. Hydrophosphite (HA) materials are the most popular in bone grafts. We would like to present a sterilization method using Q-switch Nd:YAG laser with high output energy among non-ionizing radiation suitable for small medical devices. In this study, sterilization power was most ideally shown at UV wavelengths of 1.5 W, 266 nm, and 10 pulses. Different wavelength bands; infrared and visible light; showed passive sterilization, and ultraviolet A and C showed differences in sterilization according to the pulse. In laser sterilization were differences that found according to the wavelength bands and pulses.
|
|
KEYWORD
|
|
Sterilization, Nd:YAG Laser, Non-Radiation
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|