Improved X-ray Center Beam Alignment Evaluation Method
Choi Seok-Yoon

Abstract
X-ray equipment, which is frequently used in radiology and treatment, is the most common and most used equipment in clinical practice. Equipment must provide accurate information to patients through continuous quality control. In case of manual quality control measurement, reproducibility may be poor and there may be a problem with reliability of evaluation results. In this study, an automated program was developed and attempted to measure how much the central ray between the focus of the X-ray tube and the variable aperture of the diagnostic X-ray generator used in clinical practice coincides. As a result of the experiment, it succeeded in calculating the coordinates of the two center points, and the distance between the two points was calculated in pixels and applied to the judgment and the automatic judgment value for whether the center line coincidence is within the normal angle or the abnormal angle is presented. The results of this study are considered to be very helpful in the quality control of the X-ray apparatus.
KEYWORD
Central ray, Automatic Program, X-ray equipment, Tube angle