Journal of the Korean Society of Radiology
2020 Volume.14 No. 7 p.849 ~ p.856
A Study on the Improvement of Gamma Ray Energy Spectrum Resolution through Electrical Noise Reduction of High Purity Ge Detector
Lee Sam-Yol

Abstract
In the gamma-ray energy spectrum study, nuclide analysis through energy analysis is very important. High-purity Ge detectors, which are commonly used for gamma-ray energy measurements, are commonly used because of their high energy resolution and relatively high detection efficiency. However, in order to maintain a high energy resolution, the semiconductor detector has a problem in that it is difficult to maintain the original performance if the noise generated from the surrounding environment is not effectively blocked, and the effect of the expensive device is not achieved. Therefore, in this study, ground loop isolator (NEXT-001HDGL) was used to remove the electrical noise generated from the detector. In order to test the effect of improving energy resolution, HPGe detection device newly installed in the proton accelerator KOMAC was used. In the case of gamma-ray energy 2614 keV, the energy resolution was improved from (0.16 ¡¾ 0.02) % to (0.11 ¡¾ 0.01) %, and in the case of gamma-ray energy 662 keV of 137Cs isotope, the energy resolution was improved from (0.72 ¡¾ 0.07) % to (0.27 ¡¾ 0.03) %. This result is considered to be very useful for the gamma ray spectrum study using the HPGe detection equipment of KOMAC(Korea Multi-Purpose Accelerator Complex).
KEYWORD
Energy Resolution, Electric noise, HPGe detector, Gamma-ray, Energy spectrum