Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1151820210150050755
Journal of the Korean Society of Radiology
2021 Volume.15 No. 5 p.755 ~ p.760
Calculation of Shielding Rate of Radiation Protective Equipment Using the X-ray Spectrum of IPEM Report-78
Han Dong-Hyun

Abstract
In this study, the shielding rate of major X-ray protective equipment used in the medical environment was calculated using X-ray spectrum data emitted from the diagnostic X-ray generator of The Institute of Physics and Engineering(IPEM) Report-78, and the applicability of radiation protection was investigated. Radiation shielding rates were calculated through reduction rates of air-kerma and total intensity for lead apron (0.3 mmPb), thyroid shield (0.5 mmPb), lead goggles (0.5 mmPb), and lead glass (1.8, 2.7, 3.3 mmPb) used for diagnostic X-ray protection. As a result, the shielding rate calculated as the air kerma reduction rate ranged from 96.31 to 100% at 80 kV, and 90.35 to 100% at 120 kV. In addition, the results of this calculation were well matched with the results of previous studies measuring the actual shielding rate, and it is expected that the X-ray spectrum data of IPEM Report-78 can be used for radiation protection.
KEYWORD
IPEM Report-78 , X-ray Spectrum , Radiation Protection , Shielding Rate
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)