KMID : 1151820210150070975
Journal of the Korean Society of Radiology
2021 Volume.15 No. 7 p.975 ~ p.982
Magnetic Resonance Image Analysis using MESH for High-frequency Shielding
Shin Woon-Jae

Abstract
The purpose of this study is to evaluate the effect on the phantom for magnetic resonance imaging located nearby by partially shielding RF with a mesh made thinner than hair composed of copper, black metal, and polyester using metallic materials of titanium, which are commonly used for esophageal stents and implants in the body. Magnetic resonance images according to field of view (FOV) were analyzed in the Spin Echo T1 weighted images of TR 500 ms, TE 20 ms, NEX 1, and slice thickness 5mm using a Cardiac coil of 3T Achieva X-series. Aliasing artifact did not occur in FOV 304 mm ¡¿ 304 mm, but it occurred in 250 mm ¡¿ 250 mm and 170 mm ¡¿ 170 mm. In FOV 170 mm ¡¿ 170 mm, when a mesh was not used, the SNR was measured with 78.23, and when separated by standing a mesh in the middle, it was 215.05, and when completely shielded with a mesh, the SNR was 366.44. In addition, when completely shielded with a mesh, the aliasing artifact was also removed, and signal intensities on the left, middle and right of the image were also able to obtain homogeneous images compared to the previous two cases. In conclusion, if RF is partially shielded with a mesh, aliasing artifact can be removed, and magnetic resonance images with excellent image resolution and homogeneity can be obtained using a small FOV.
KEYWORD
Magnetic Resonance Imaging(MRI) , Mesh , Phantom , FOV , Aliasing artifacts