KMID : 1151820220160020131
Journal of the Korean Society of Radiology
2022 Volume.16 No. 2 p.131 ~ p.139
Image Evaluation Analysis of CT Examination for Pedicle Screw Insertion
Hwang Hyung-Suk

Im In-Chul
Abstract
The purpose of this study was to insert a pedicle screw into a pig thoracic vertebrae, a general CT scan(Non MAR), and a thoracic axial image obtained with the Metallic Artifact Reduction for Orthopedic Implants (O-MAR) to reduce artifacts. The image obtained by reconstructing the algorithm (Standard, Soft, Bone, Detail) was used using the image J program. Signal to noise ratio(SNR) and contrast to noise ratio(CNR) were compared and analyzed by obtaining measured values based on the given equation. And this study was to investigate tube voltage and algorithm suitable for CT scan for thoracic pedicle screw insertion. As a result, when non-MAR was used, the soft algorithm showed the highest SNR and CNR at 80, 100, 120, and 140 kVp, On the other hand, when MAR was used, the standard algorithm showed the highest at 80 kVp, and the standard and soft algorithms showed similar values at 100 kVp. At 120 kVp, the Soft and Standard algorithms showed similar values, and at 140 kVp, the Soft algorithm showed the highest SNR and CNR. Therefore, when comparing Non-MAR and MAR, even if MAR was used, SNR and CNR did not increase in all algorithms according to the change in tube voltage. In conclusion, it is judged that it is advantageous to use the Soft algorithm at 80, 100, 120, and 140 kVp in Non MAR, the Standard algorithm at 80 and 100 kVp in MAR, and the Soft algorithm at 120 and 140 kVp. This study is expected to serve as an opportunity to further improve the quality of images by using selective tube voltage and algorithms as basic data to help evaluate images of pedicle screw CT scans in the future.
KEYWORD
Pedicle screw, Artifact, Metallic Artifact Reduction for Orthopedic Implants(O-MAR), Algorithm
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)