Journal of the Korean Society of Radiology
2022 Volume.16 No. 2 p.151 ~ p.156
Digital Position Measurement with MLPE of PET detector using a Small Number of Photosensors
Kang Seung-Hun

Lee Seung-Jae
Abstract
A detector using a small number of photosensors was designed, and the position of a scintillation pixel that interacted with gamma rays through a maximum likelihood position estimation(MLPE) was measured as a digital position. For this purpose, simulation was performed using DETECT2000, which can simulate the movement of light within the scintillator, and the accuracy of position measurement was evaluated. A detector was configured using a 6 6 scintillation pixel array and 4 photosensors, and a gamma ray event was generated at the center of each scintillation pixel to create a look-up table through the ratio of acquired light. The gamma-ray event generated at the new position was applied as the input value of the MLPE, and the positiion of the scintillation pixel was converted into a digital positiion after comparison with the look-up table. All scintillation pixels were evaluated, and as a result, a high accuracy of 99.1% was obtained. When this method is applied to the currently usesd system, it is concidered that the process of determining the position of the scintillation pixel will be simplified.
KEYWORD
Digital position, MLPE, photosensor, DETECT2000