Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1155520140090040277
Anesthesia and Pain Medicine
2014 Volume.9 No. 4 p.277 ~ p.281
Efficacy of noninvasive pulse co-oximetry as compared to invasive laboratory-based hemoglobin measurement during spinal anesthesia
Chung Jin-Hun

Ji Jae-Young
Kim Nan-Seol
Seo Yong-Han
Gong Hyung-Youn
Kim Jae-Woo
Kim Jong-Bun
Yoo Sie-Hyeon
Abstract
Background: The Masimo Radical 7 (Masimo Corp., Irvine, CA, USA) pulse co-oximeter¨Þ noninvasively determines the hemoglobin concentration using the principle of transcutaneous spectrophotometry. We compared hemoglobin levels determined using this device (SpHb) with those determined using an invasive laboratory-based technique (tHb) during spinal anesthesia.

Methods: Thirty patients received spinal anesthesia with 0.5% hyperbaric bupivacaine. The pulse co-oximeter probe was mounted on the second toe, and arterial blood samples were obtained from a radial artery catheter. SpHb, tHb, and perfusion index (PI) values were recorded before and 20 and 40 min after intrathecal injection of bupivacaine.

Results: Before spinal anesthesia, the SpHb and tHb showed a significant difference of ?2.86 ¡¾ 1.56 g/dl (P £¼ 0.005), but no significant differences were found between tHb and SpHb at 20 and 40 min after spinal anesthesia (?0.16 ¡¾ 2.45 g/dl and 0.29 ¡¾ 2.68 g/dl). Additionally, PI was significantly increased at 20 and 40 min after spinal anesthesia compared to the pre-anesthetic value (P £¼ 0.001).

Conclusions: The toe is not the monitoring site for pulse co-oximetry in adult patients, but the pulse co-oximetry on the toe appears to be appropriate as a noninvasive hemoglobin monitoring device after spinal anesthesia. (Anesth Pain Med 2014; 9: 277-281)
KEYWORD
Hemoglobin, Pulse co-oximetry, Spinal anesthesia
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø