Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1155520240190010035
Anesthesia and Pain Medicine
2024 Volume.19 No. 1 p.35 ~ p.43
Application of the Bair Hugger¢â core body temperature at wrist region with upper body warming blanket: a prospective observational study
Oh Kyung-Seo

Park Yong-Hee
Baek Chong-Wha
Woo Young-Cheol
Abstract
Background: Body temperature monitoring is essential during the perioperative period. However, core body temperature measurement requires invasive device that may cause complications. This study aimed to evaluate the accuracy of non-invasive Bair Hugger¢â core body temperature monitoring system (BHTMS) at the wrist compared with esophageal temperature under general anesthesia

Methods: Twenty adult patients of the American Society of Anesthesiologists physical status I or II were enrolled. BHTMS sensor was applied at wrist region. After tracheal intubation, an esophageal probe was inserted. Bair Hugger¢â upper body warming blankets were used. Esophageal temperature (Teso) and BHTMS at wrist (Twrist) were recorded every 10 min.

Results: Total of 257 pairs of data sets were analyzed: Teso and Twrist had no statistically significant difference (P = 0.103). Median of Teso and Twrist were 36.5¡ÆC and 36.4¡ÆC. Bland-Altman analysis showed Teso ? Twrist of 0.14¡ÆC ¡¾ 1.44. Subsequently, 99 pairs of 0?40 min data set were analyzed and showed significant difference at 0 and 10 min (P < 0.001) but no significant difference at 20, 30 and 40 min. Bland? Altman plot by times showed difference (Teso - Twrist) of 1.49¡ÆC ¡¾ 2.00, 0.82¡ÆC ¡¾ 1.30, 0.29¡ÆC ¡¾ 1.32, ?0.03¡ÆC ¡¾ 0.84, and ?0.12¡ÆC ¡¾ 0.82 at 0, 10, 20, 30 and 40 min respectively.

Conclusions: BHTMS at wrist area under the upper body warming blanket is a potential alternative other than esophageal temperature for monitoring body temperature after 30 min of anesthesia induction.
KEYWORD
Body temperature, Hypothermia, Intraoperative care, Intraoperative monitoring, Perioperative period, Thermometers
FullTexts / Linksout information
Listed journal information