Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1159320080100010021
Korean Journal of Vision Science
2008 Volume.10 No. 1 p.21 ~ p.31
The Occurrence of Spherical Equivalent Error Between Subjective and Objective Refraction According to Cylindrical Refraction Error After Wearing Contact Lenses
Lee Dong-Yeol

Abstract
Purpose & Methods: In order to compare the occurrence of spherical equivalent error between subjective refraction and objective refraction(using autorefractometer) according to cylindrical refractive error, over-refraction was performed from contact lens wearers.

Results: The differences of measured values by objective refraction in contrast to subjective refraction on wearing 5 types contact lens, in cases of one-day soft contact lens, 6-month contact lens and silicon-hydrogel contact lens, there was the biggest error of spherical equivalent in group B(-0.25~-0.75D of cylindrical refractive error). In cosmetic color contact lens, there was the biggest error in group D(up to -2.00D). In RGP contact lens, there was the biggest error in group A(0.00D). But their differences were no more than 0.10D, it was not significant clinically.

Conslusion: From these results, autorefractometer should be an useful objective refraction in actual affairs.
KEYWORD
Over-refraction, Subjective refraction, Objective refraction, Spherical equivalent
FullTexts / Linksout information