Journal of Korean Society for Radiotherapeutic Technology
1992 Volume.5 No. 1 p.115 ~ p.119



Abstract
The beam characteristics and dosimetric measurements of the 6MV X-ray and 6MeV electron beam from a ML-6M linear accelerator are examined. The Percent Depth Dose(PDD) table and the tissue Maximum Ratio(TMR) table are taken from measurement as a function of the field size and the depth. The calculated TMR table from PDD table is compared with those from measurement. Other beam characteristics such as output factor, beam profile(including flatness, symmetry and penumbra), wedge, and the variation of Dmax are presented. All of these dosimetric measurements sufficiently characterized the beam to permit safe clinical use.