KMID : 1164219940060010089
|
|
Journal of Korean Society for Radiotherapeutic Technology 1994 Volume.6 No. 1 p.89 ~ p.93
|
|
Determination of the Virtual Focus Position for Electron Beam with Air Scanning
|
|
Kwon Kyung-Tae
Park Kwang-Ho Kim Jung-Man
|
|
Abstract
|
|
|
Authors have measured virtual source distance of electron beam from CL/1800 medical linear accelerator, with newly designed method. Beam scanning was performed with the direction of beam axis in the air. Compared results between this study and well established in phantom measurement shows good agreement with in experimental error. And we have found that build-up cap plays very important role in air measurement because of charge build up. The method of in-air measurement of virtual source distance is very easy to set-up and generate accurate results.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|