KMID : 1164219980100010088
|
|
Journal of Korean Society for Radiotherapeutic Technology 1998 Volume.10 No. 1 p.88 ~ p.93
|
|
A Study on the Improvement of Penumbra and Dose Distribution in the Multileaf Collimator Field Edge
|
|
Kim Chang-Wook
Na Soo-Kyung Park Byung-Sub
|
|
Abstract
|
|
|
Multileaf collimator is essential equipment in conformal radiation therapy, however the use is limitted by increase of penumbra width and undulating dose distribution at the field edge. The purpose of this study is to improve the penumbra and dose distribution in the multileaf collimator field edge. Measurement were performed with X-omat V film in solid water phantom using 6MV photon beam from Siemens linear accelerator. All the measurement were made along the central axis of circular field for constant SSD of 100 cm. To improve the penumbra and dose distribution collimator was rotated by 15 degrees from 0 to 90 degrees (collimator rotation method) and center was shifted to the longitudinal direction by fourth of lead width (center shift method). We compare the penumbra and dose distribution at the field edge to alloy block. Dose distribution and penumbra width at the feild edge of MLC showed undulated dose pattern and increased penumbra compared with alloy block. However, in the collimator rotation method and center shift method we abtained simular results with alloy block. Through the study we expected that clinical use of MLC will be increase.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|