Journal of Korean Society for Radiotherapeutic Technology
2000 Volume.12 No. 1 p.91 ~ p.104
Study on 6 MV Photon beam Dosimetry by Asymmetric Collimator Variation of Linear Accelerator
Yoon Joo-Ho

Lee Chul-Soo
Abstract
Recently linear accelerator in radiation therapy in asymmetric field has been easily used since the improvement and capability of asymmetrical field adjustment attached to the machine. It has been thought there have been some significant errors in dose calculation when asymmetrical radiation fields have been utilized in practice of radiation treatments if the fundamental data for dose calculation have been measured in symmetrical standard fields. This study investigated how much the measured data of dose distributions and their isodose curves are different between in asymmetrical and symmetrical standard fields, and how much there difference affect the error in dose calculation in conventional method measured in symmetrical standard field. The distributions of radiation dose were measured by photon diode detector in the water phantom (RFA-300P, Scanditronix, Sweden) as tissue equivalent material on utilization of 6 MV linear accelerator with source surface distance (SSD) 1000 mm. The photon diode detector has the velocity of 1 mm per second from water surface to 250 mm depth in the field size of asymmetrical fields. The measurements of percent depth dose (PDD) and subsequent plotting of their isodose curves were performed from water surface to 250mm dmm from Y-center axis in field in order to absence the variability of depth dose according to increasing field sizes and their affects to plotted isodose curves. The difference of PDD between symmetric and asymmetric field was maximum , the larger decrease difference of PDD as the greater field size and as greater the depth, The difference of PDD between asymmetrical field and equivalent square field showed maximum , and the larger decreased differenced PDD as the greater field size and as greater the depth, these differences of PDD were out of of dose calculation as defined by international Commission on radiation unit and Measurements(ICRU). In the dose distribution of asymmetrical field (half beam) the plotted isodose curves were observed to have deviations by decreased PDD as greater as the blocking of the beam moved closer to the central axis, and as the asymmetrical field increased by moving the block 10 mm keeping away from the central axis, the PDD increased and plotted isodose curves were gradually more flattened, due to reduced amount of the primary beam and the fraction of low energy soft radiations by passing thougepth in asymmetrical field by moving independent jaw each 10 h beam flattening filter. As asymmetrical radiation field as half beam radiation technique is used, the radiation dosimetry calculated in utilizing the fundamental data which measured in standard symmetrical field should be converted on bases of nearly measured data in asymmetrical field, measured beam data flies of various asymmetrical field in various energy and be necessary in each institution.
KEYWORD
Asymmetric field, Dose distribution, 6MV linear accelerator
FullTexts / Linksout information
 
Listed journal information