Precision evaluation of the treatment that used coordinates confirmation of couch in case of two forgets adjoined
Seo Jeong-Min

Song Ki-Won
Park Young-Hwan
Jung Chun-Young
Abstract
Purpose : Confirming an error to be able to break out in a method to move couch manually while operator sees the skin marks on patient in case of curing head who got 2 targets adjoined, so we analyze coordinates price of couch, evaluate reproducibility and precision of change movements between targets.

Materials and Methods : In radiotherapy, for confirming errors in manual movements by operators by exchanging between two targets to treat patient head, we read coordinates price(vertical, longitudinal, lateral three directions of couch) shown on a monitor of LINAC( CL 2100, Varian, USA) in order to evaluate accuracy about the length that moved in time for moving couch manually. After reading movement length of coordinates recorded in three directions of all treatment, we compared distance between targets recorded in RTP(Pinnacle, ADAC, USA) with reading coordinates price of couch, setting actually done the same patient for ten times, coordinates were recorded, treated for evaluating averages and degrees of errors and standard deviations.

Results : In method to confirm skin marks of patient by operators¡¯ view and to move couch manually, average standard deviations of movements between two targets are vertical 1.4mm, longitudinal 0.9mm, lateral 2.2mm in each direction. As for the error in straight dimension, it is about 3.6mm averages and 5.1mm maximum. The average of errors in each directions was vertical 1mm, longitudinal 0.7mm, lateral 2.7mm. The greatest error broke out in lateral direction with of all cases ; to exceed an error average.

Conclusions : If operators moved manually couch for changing target points, errors about 3.6mm average degrees occur. It is important that operators confirm the errors prices of actual couch coordinates for asking a correct movement between the targets adjoined each other ; in case of treatment demanding high precision like 3D conformal therapy or IMRT. Therefore, if we apply couch coordinates confirmation to reproducibility and to precision evaluation of treatment, it¡¯s expected that we can execute high-quality radiotherapy.