KMID : 1164220040160020081
|
|
Journal of Korean Society for Radiotherapeutic Technology 2004 Volume.16 No. 2 p.81 ~ p.89
|
|
The dose distribution and DVH change analysis wing to effect of the patient setup error
|
|
Kim Gyeong-Tea
Ju Sang-Gyu Ahn Jae-Hong Park Young-Hwan
|
|
Abstract
|
|
|
Introduction : The setup error due to the patient and the staff from radiation treatment as the reason which is important the treatment record could be decided is a possibility of effect. The SET-UP ERROR of the patient analyzes the effect of dose distribution and DVH from radiation treatment of the patient.
Material & Methode : This test uses human phantom and when C-T scan doing, It rotated the Left direction of the human phantom and it made SET-UP ERROR , Standard plan and 3mm, 5mm, 7mm, 10mm, 15mm, 20mm with to distinguish, it made the C-T scan error. With the result, The SET-UP ERROR got each C-T image Using RTP equipment It used the plan which is used generally from clinical - Box plan, 3Dimension plan( identical angle 5beam plan) Also, ( CTV+1cm margin, CTV+0.5cm margin, CTV+0.3,cm margin = PTV) it distinguished the standard plan and each set-up error plan and The plan used a dose distribution and the DVH and it analyzed
Result : The Box4 the plan and 3Dimension plan which it bites it got similar an dose distribution and DVH in 3mm, 5mm From rotation error and Rectilinear movement( ). Rotation error and rectilinear error 7mm, 10mm, 15mm, 20mm appeared effect it will go mad to a enough change in treatment ( )
Conclusion : The diminishes the effect of the SET-UP ERROR must reduce move with tension of the patient Also, we are important accessory development and the supply that it reducing of reproducibility and the move
|
|
KEYWORD
|
|
DVH, dose distribution, setup error
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|