Journal of Korean Society for Radiotherapeutic Technology
2005 Volume.17 No. 1 p.9 ~ p.17
A comparative study of dose distribution for whole brain with field-in-field technique
Kim Bo-Gyum

Lee Je-Hee
Jung Chi-Hoon
Park Heung-Deuk
Abstract
Purpose : Uniform dose distribution of the target volume is very important in the radiation treatment. We will evaluate the usefulness of Field-in-Field Technique use to get uniform dose distribution of the target volume and try to find Apply possibility out to a whole brain treatment patient of various thickness.

Material and method : We compare the dose distribution when we applied Field-in-Field Technique and parallel opposed fields technique. establish the treatment plan to a phantom(acryl 16cm spheral phantom) and do the measurement, assessment use the TLD and Low sensitivity film. Also the assessment did Apply possibility of Field-in-Field Technique to 20 patient object of various thickness.

Result : In the case to use the parallel opposed fields at the whole brain treatment high dose region appeared but reduce to lesses when we used the Field-in-Field technique. We could get similar numerical value the film and TLD measurement result also. The change of the dose distribution appeared to its although it applied such Field-in-Field technique to various patient so that we were identical.

Conclusion : We can get uniform dose distribution of in the treatment region if we apply the Field-in-Field technique at the whole brain treatment. Also alternate can play the role of the wedge filter and 3D compensator and We are thought by minimizing the obstacle to be happened due to the high dose region when radiation treatment.
KEYWORD
field-in-field, whole brain treatment, dose distribution