KMID : 1164220050170010033
Journal of Korean Society for Radiotherapeutic Technology
2005 Volume.17 No. 1 p.33 ~ p.40
The evaluation of custo-made EWHA Breast Device(EBD)
Byun Young-Sik

Oh Tae-Sung
Park Chong-Yil
Shin Hyun-Kyoh
Abstract
Purpose : We recently perform the latest radiotheraphy (3D conformal, IMRT,etc.) with the development of 3D CT planning technology. However, in case patients with breast cancer, it is difficult to acquire the CT image with the limitation of CT hole size and tilting of breast immobilization device. The Ewha Breast Device(EBD) was constructed to improve the problem in the treatment of patients with breast cancer and we are intend to introduce the procedure of the EBD construction in this study and compare the EBD with conventional breast device in the view point of usefulness.

Materials and Methods : We have constructed the EBD with acryl, analyze the skull size with CT data, consider the skin folder in SCL field and evaluated the EBD usefulness from the view point of set-up reproducibility, dose distribution, skin reaction in comparison with conventional breast device.

Results : In the case of patients set-up error analysis, the EBD is superior to conventional device in portal film repetition() check (80pt.), equal to that in simulation & CT image coincidence check(5pt.). There is no difference between the two systems in dose distribution and skin reaction in SCL field is better the EBD than conventional device.

Conclusions : The construction of the EBD enable us to perform the latest radiotheraphy in breast treatments, relieve the pains in simulation, and reduce, the skin reaction. In the future, we expect that modification of the EBD is useful in treating for patients with breast cancer.
KEYWORD
EBD(EWHA Breast Device), skin reaction