KMID : 1164220050170010041
|
|
Journal of Korean Society for Radiotherapeutic Technology 2005 Volume.17 No. 1 p.41 ~ p.43
|
|
Quality assurance for computed-tomography simulator : Report of the AAPM Radiation Therapy Committee Task Group No.66
|
|
Lee Yun-Seok
|
|
Abstract
|
|
|
Purpose : Wish to present degree management process that is efficient confirm radiation treatment exclusive use CT simulator¡¯s Q.A item that become Q.A and Differentiation of diagnosis area that present Report of the AAPM Task Group No.66 using Q.A tool that produce itself and secure safe and correct CT-simulation process and equip convenience.
Method and material : Manufacture CT simulator¡¯s Q.A tool on source and confirm virtue between isocenter of wall laser system, patient table, CT scanner¡¯s imaging plane that present in Report of the AAPM Task Group No.66 by daily publication unit.
Result : Confirmed measured value from Report of the AAPM Task Group No.66 to confirmation of presenting degree management item in wall laser¡¯s , table¡¯s , imaging plane¡¯s tolerance extent.
Conclusion : There is unconfirmed item from CT-simulation process for therapy to CT Q.A protocol of existent diagnosis area, premising suitable degree management of radiation treatment exclusive use CT-simulator equipment confirming presenting Q.A item in Report of the AAPM Task Group No.66 safe and correct CT-simulation process guarantee can
|
|
KEYWORD
|
|
Qualify assurance, QA tool
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|