KMID : 1164220050170020105
Journal of Korean Society for Radiotherapeutic Technology
2005 Volume.17 No. 2 p.105 ~ p.111
Manufacture of Daily Check Device and Efficiency Evaluation for Daily Q.A
Kim Chan-Yong

Jae Young-Wan
Park Heung-Deuk
Lee Je-Hee
Abstract
Purpose : Daily Q.A is the important step which must be preceded in a radiation treatment. Specially, radiation output measurement and laser alignment, SSD indicator related to a patient set-up recurrence must be confirmed for a reasonable radiation treatment. Daily Q.A proceeds correctness and a prompt way, and needs an objective measurement basis. Manufacture of the device which can facilitate confirmation of output measurement and appliances check at one time was requested.

Materials and Methods : Produced the phantom formal daily check device which can confirm a lot of appliances check (output measurement and laser alignment. field size, SSD indicator) with one time of set up at a time, and measurement observed a linear accelerator (4 machine) for four months and evaluated efficiency.

Results : We were able to confirm an laser alignment, field size, SSD indicator check at the same time, and out put measurement was possible with the same set up, so daily Q.A time was reduced, and we were able to confirm an objective basis about each item measurement. As a result of having measured for four months, output measurement within , and measured laser alignment, field size, SSD indicator in range within .

Conclusion : We can enforce output measurement and appliances check conveniently, and time was reduced and was able to raise efficiency of business. We were able to bring a cost reduction by substitution expensive commercialized equipment. Further It is necessary to makes a product as strong and slight materials, and improve convenience of use.
KEYWORD
linear accelerator, daily Q.A
FullTexts / Linksout information
 
Listed journal information