Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1164220050170020133
Journal of Korean Society for Radiotherapeutic Technology
2005 Volume.17 No. 2 p.133 ~ p.140
Evaluation of Usefulness on In-vivo Diode Dosimetry for Measuring the Tumor Dose of Oral Cancer Patient
Na Kyoung-Su

Lee Je-Hee
Park Heung-Deuk
Abstract
Purpose : This test is designed to identify the validity of treatment plan by implementing real-time dosimetry by means of dose that is absorbed into PTV and OAR when preparing doses of 3D and POP plans.

Materials and Methods : In treatment. error can be calculated be comparing Exp. Dose with the actual dose, which has been converted from ¡¯the reading value obtained by placing diode detector on the area to be measured¡¯. Same test can be repeated using Alderson-Rando phantom.

Results : Errors were found: A patient(POP plan): 197.6/199=-1.2%, B patient(3D-plan): 199.9/198.7=+0.6%, C patient: 196/200=-1.5%. In addition, considering the resulted value of measuring OAR besides target-dose for C patient showed 96/200, representing does of 47%, the purpose of protection was judged to be duly accomplished. Also it was acknowledged the resulted value of -3.7% met the targeted dose within the range of .

Conclusion : Aimed for identifying the usefulness of pre-treatment dose measurement using diode detector, this test was useful to evaluate the validity of curing because it resulted in the identification of category to be protected as well as t dose. Moreover, it is thought to have great advantage in ascertaining the dose of target, dose of which is not calculated yet. Similar to L-gram before treatment, this test is thought to be very effective so that it can bring great advantages in the aspects such as validity of curing method and post-treatment plan as well.
KEYWORD
in-vivo diode dosimetry, diode
FullTexts / Linksout information
 
Listed journal information