KMID : 1164220060180020119
Journal of Korean Society for Radiotherapeutic Technology
2006 Volume.18 No. 2 p.119 ~ p.125
Effectiveness of a Custom-made Multi Purpose Brachytherapy Phantom
Jang In-Gi

Lee Joung-Jin
Kim Wan-Sun
Abstract
Purpose: To measure the dose for dose optimization at the reference point (A, B) and the critical organ with multi Purpose brachytherapy phantom (MPBP). For this wort the MPBP was custom made, and designed to reconstruct the treatment applicator using multi function applicator (MFA) in the same way as the treatment of patient.

Materials and Methods: Dose measurements were made at the reference points (A, B) and the bladder with thermoluminescence dosimeter (TLD) for four patients with tandem and ovoid of uterine cervix cancer using the phantom. In Phantom, Total 20 times of the measurements were made with 5 times a patient.

Results: The results of TLD measurements in MPBP phantom showed the relative error ranging from -3.2% to 3.8% at A point, and -1.4% to 4% at B point and 1.3% to 7.15% at the bladder of reference point.

Conclusion: The reproducibility of dose measurement under the same condition as the treatment could be achieved using the custom-made MFA in phantom and the dose at the reference point (A, B) and bladder could be analyzed accurately. The measured dose acquired in MPBP can apply for the dose optimization.
KEYWORD
brachytherapy phantom, purpose brachytherapy phantom(MPBP), multi function applicator(MFA), thermoluminescence dosimeter(TLD)