Journal of Korean Society for Radiotherapeutic Technology
2007 Volume.19 No. 1 p.43 ~ p.49
Implementation of KV Cone Beam CT for Image Guided Radiation Therapy
Yoo Young-Seung

Lee Hwa-Jung
Kim Dae-Young
Yu Ri
Abstract
Purpose: The aim of this study was the clinical implementation of IGRT using KV CBCT for setup correction in radiation therapy.

Materials and Methods: We selected 9 patients (3 patient for each region; head, body, pelvis)and acquired 135 CBCT images with CLINAC iX (Varian medical system, USA). During the scan, the required time was measured. We analyzed the result in 3 direction; vertical, longitudinal, lateral.

Results: The mean setup errors at the couch position of vertical, lateral, and longitudinal direction were 0.07, 0.12, and 0.1 cm in the head region, 0.3, 0.26, and 0.22 cm in the body region, 0.21, 0.18, and 0.15 cm in the pelvis region respectively. The mean time required for CBCT was 6~7 minute.

Conclusion: The CBCT on the LINAC provides the capacity for soft tissue imaging in the treatment position and real time monitoring during treatment delivery. With presented workflow, the setup correction within reasonable time for more accurate radiation therapy is possible. And it¡¯s image can be very useful for adaptive radiation therapy(ART) in the future with improved image quality.
KEYWORD
IGRT, CBCT, set up error