A Study on Prospective Plan Comparison using DVH-index in Tomotherapy Planning
Kim Joo-Ho

Cho Jung-Heui
Lee Sang-Kyoo
Jeon Byeong-Chul
Yoon Jong-Won
Kim Dong-Wook
Abstract
Purpose: We proposed the method using dose-volume Histogram index to compare prospective plan trials in tomotherapy planning optimization.

Materials and Methods: For 3 patients in cranial region, thorax and abdominal region, we acquired computed tomography images with PQ 5000 in each case. Then we delineated target structure and normal organ contour with pinnacle Ver 7.6c, after transferred each data to tomotherapy planning system (hi-art system Ver 2.0), we optimized 3 plan trials in each case that used differ from beam width, pitch, importance. We analyzed 3 plan trials in each region with isodose distribution, dose-volume histogram and dose statistics. Also we verified 3 plan trials with specialized DVH-indexes that is dose homogeneity index in target organ, conformity index around target structure and dose gradient index in non-target structures.

Results: We compared with the similarity of results that the one is decide the best plan trial using isodose distribution, dose volume histogram and dose statistics, and the another is using DVH-indexes. They all decided the same plan trial to better result in each case.

Conclusion: In some of case, it was appeared a little difference of results that used to DVH-index for comparison of plan trial in tomotherapy by special goal in it. But because DVH-index represented both dose distribution in target structure and high dose risk about normal tissue, it will be reasonable method for comparison of many plan trials before the tomotherapy treatments.
KEYWORD
tomotherapy, DVH-index, plan comparing