Journal of Korean Society for Radiotherapeutic Technology
2008 Volume.20 No. 2 p.103 ~ p.107
Surface Dose Measurement of Electron Beam within the Magnetic Field Variation
Je Jae-Yong

Noh Kyung-Suk
Shin Woon-Jae
Park Cheol-Woo
Abstract
Purpose: This paper describes a electron field presence of magnetic field, intensity and shape surface dose variation to clinical application possibility.

Materials and Methods: The using 6 MeV electron and 10¡¿10§² field size, 9 hole to shielding block make the by measure the film, when the magnetic field position inside and outside of the X-Omat film and parallel plate ionization chamber using the surface dose measured.

Results: Present of 4 cm to the side at angle about 3 degree from beam center, use of ring type magnetic is 0.9% increase the surface dose, lens block located in the magnetic field the surface dose 1.58% increase, half magnetic field¡¯s position on the side of them at the field center of the 3.6% increase of the surface dose.

Conclusion: Surface dose variation is with magnetic field about the mean electron beam of progress direction change, orbit region patient¡¯s is inconvenient without surface dose increase percentage case goodness will be used as a useful way.
KEYWORD
electron beam, magnetic field, surface dose