Consideration about Ozone Generation in the Treatment Room While Treating a Patient
Kwak Yong-Kuk

Yoon Il-Kyu
Lee Je-Hee
Yoo Suk-Hyun
Abstract
Purpose: Measure the ozone level in the treatment room while treating a patient so want to know the degree of contamination caused by ozone occurrence.

Materials and Methods: Use the linear accelerator (Clinac 21EX, Varian, USA) with the ozone meter (series-200, aeroQual, New Zealand) and water phantom (Wellhofer, IBA, Germany) is irradiated the radiation so that measured the ozone generation level according to MU, dose-rate, SSD, field size, energy, delay time and put the ozone meter in the treatment room actually while treating a patient so measured the daily ozone level variation.

Results: While irradiating the radiation, degree of ozone contamination wasn¡¯t affected by the energy but mostly in case of electron beam, ozone level was higher than photon beam. The higher dose-rate (0.016~0.025 ppm/hr), the farther SSD (0.018~0.030 ppm/hr), the wider field sizes (0.016~0.025 ppm/hr), the more MU (0.018~0.046 ppm/hr), it occurred high ozone level. Ozone decrement according to delay time changed the background level (0.016 ppm/hr) after elapsed time of 10 minutes from irradiating radiation. And daily ozone occurrence level in the treatment room was below ozone standard level 0.1 ppm/hr (average:0.06 ppm/8 hr) but it could confirm that ozone generation level was included the level (max:0.038 ppm/hr) above 0.02 ppm/hr which patient could perceive.

Conclusion: Through ozone level according to variation of certain conditions, actually in the treatment room ozone generation level didn¡¯t damaged to patients or workers. Commonly peoples think that ozone was harmful gas but it thought that small amount of ozone generation level while treating a patient was beneficial in the treatment room through air purge action of pathogenic germ or virus sterilization.
KEYWORD
ozone, treatment room, level, linear accelerator, ozone meter