Journal of Korean Society for Radiotherapeutic Technology
2010 Volume.22 No. 1 p.53 ~ p.60
The Study on Properties and Application of Enhanced Dynamic Wedge Factor
Kim Dae-Sup

Ban Tae-Joon
Yeom Mi-Suk
Yoo Soon-Mi
Lee Woo-Suk
Back Geum-Mun
Kwon Kyung-Tae
Abstract
Purpose: We try to calculate EDW-factor easily with the formula applies essential data of EDW-factor and evaluate the validity through a measurement.

Materials and Methods: We used the given value of GSTT (Golden Segmented Treatment Table) for the calculation of the EDW-factor. As to the experimental device, 0.6 cc farmer-type ion-chamber, an electrometer and water- phantom were used. A measurement was made at the maximum dose depth of the photon beam energy 6 MV and 15 MV under the condition that SSD (Source to Surface Distance) was 100 cm. The angle of the EDW (Enhanced Dynamic Wedge) which we use in an experiment was 60 degree, 30 degree, 20 degree in the Y1-OUT direction. We used Eclipse planning system (Varian, USA) as RTP system and the EDW-factor was calculated about all fields and EDW direction. In order to show the EDW-factor feature, a measurement was made at the selected field that verify the influence of the dependability about X, Y jaw and off-axis field.

Results: When we change the Y1 field, it influence on the EDW-Factor and measured value. But the error between measured values and calculated values was less than 1%. The experimental result indicated the tendency that the error of the result of calculation and measured value becomes smaller as the EDW angle become smaller whether the calculation point (measurement point) and iso-center are same or not. The influence of the field size and energy did not show up. We simulated with the same condition using the RTP system. And we found that it makes no difference between the MU which is calculated manually by applying the EDW-Factor obtained from the commercial program and the value which is calculated by using RTP system.

Conclusion: We excluded fitting value from well-known EDW-Factor formula and calculated EDW-factor with the formula applies essential data of EDW-factor only. As a result, there are no significant difference between the measured value and calculated value and it showed errors less than 1%. Also, we implemented the commercial program to calculate EDW-Factor conveniently without measure a factor on each field.
KEYWORD
wedge factor, dynamic wedge, enhanced dynamic wedge, enhanced dynamic wedge factor