Journal of Korean Society for Radiotherapeutic Technology
2010 Volume.22 No. 2 p.113 ~ p.122
Accuracy Evaluation of Tumor Therapy during Respiratory Gated Radiation Therapy
Jang Eun-Sung

Kang Soo-Man
Lee Chul-Soo
Kang Se-Sik
Abstract
Purpose: To evaluate the accuracy of a target position at static and dynamic state by using Dynamic phantom for the difference between tumor¡¯s actual movement during respiratory gated radiation therapy and skin movement measured by RPM (Real-time Position Management).

Materials and Methods: It self-produced Dynamic phantom that moves two-dimensionally to measure a tumor moved by breath. After putting marker block on dynamic phantom, it analyzed the amplitude and status change depending on respiratory time setup in advance by using RPM. It places marker block on dynamic phantom based on this result, inserts Gafchromic EBT film into the target, and investigates 5 Gy respectively at static and dynamic state. And it scanned investigated Gafchromic EBT film and analyzed dose distribution by using automatic calculation.

Results: As a result of an analysis of Gafchromic EBT film¡¯s radiation amount at static and dynamic state, it could be known that dose distribution involving 90% is distributed within margin of error of 3 mm.

Conclusion: As a result of an analysis of dose distribution¡¯s change depending on patient¡¯s respiratory cycle during respiratory gated radiation therapy, it is expected that the treatment would be possible within recommended margin of error at ICRP 60.
KEYWORD
respiratory gated radiation therapy, dynamic phantom, 4D-CT, RPM (real-time position management)