Journal of Korean Society for Radiotherapeutic Technology
2010 Volume.22 No. 2 p.131 ~ p.134
In Tomotherapy, for the Maintenance of Body Temperature due to Substance Use, Changes in Dose Assessment in the Organization
Hwang Jae-Woong

Jung Do-Hyeung
Kim Dae-Woong
Yang Jin-Ho
Choi Gye-Sook
Abstract
Purpose: TOMO therapy treatment for a relatively long run Beam time and temperature-sensitive detector, such as CT clinics in optimal temperature (2021¡Æ) to maintain a constant temperature in addition to its own Chamber Cooling system is activating. TOMO This clinic has been reduced in the patients¡¯ body temperature to keep the sheets and covers over the treated area. Therefore, these materials for any changes in the organization gives the dose were analyzed.

Materials and Methods: To compare changes in the organization Dose Phantom cheese (Cheese Phantom) were used, CT-simulation taking the center point of the cheese phantom PTV (Planning Target Volume, treatment planning target volume) by setting Daily dose 200 cGy, 3 meetings planned treatment. PTV, PTV +7 cm, PTV +14 cm, the total count points on the phantom using the Ion chamber cover without any substance to measure the dose, and one of the most commonly used treatment, including the frequently used four kinds of bedding materials (febric 0.8 mm, gown 1.4 mm, rug, 3.3 mm, blanket 13.7 mm) and covered with a phantom and the dose measured at the same location were analyzed 3 times each.

Results: PTV, PTV +7 cm, PTV +14 cm from the point of any substance measured in the state are covered with four kinds of materials (fabric, gown, rug, blanket) was measured in the covered states and compares their results, PTV respectively -0.17%, -0.44%, -0.53% and -0.9% change, PTV +7 cm, respectively -0.04%, +0.07%, +0.06%, +0.07%, were changed, PTV +14 cm, respectively 0%, -0.06%, -0.02%, +0.6%, respectively.

Conclusion: These results TOMO treatment to patients to maintain their body mass by using PTV thickness of the material decreased in proportion to. PTV +7 cm, but showed slight changes in the point, PTV +14 cm at the point of the dose was increased a little. Sejijeom all the difference in treatment tolerance ¡¾3% range, this is confirmed in the coming treatment will not affect the larger should be considered.
KEYWORD
tomotherapy, PTV(planning target volume)