KMID : 1164220110230010001
Journal of Korean Society for Radiotherapeutic Technology
2011 Volume.23 No. 1 p.1 ~ p.6
Utility Estimation of the Manufactured Stereotactic Body Radiotherapy Immobilization
Lee Dong-Hoon

Ahn Jong-Ho
Seo Jeong-Min
Shin Eun-Hyuk
Choi Byeong-Ki
Song Gi-Won
Abstract
Purpose: Immobilizations used in order to maintain the reproducibility of a patient set-up and the stable posture for a long period are important more than anything else for the accurate treatment when the stereotactic body radiotherapy is underway. So the purpose of this study is to adapt the optimum immobilizations for the stereotactic body radiotherapy by comparing two commercial immobilizations with the self-manufactured immobilizations.

Materials and Methods: Five people were selected for the experiment and three different immobilizations (A: Wing-board, B: BodyFix system, C: Arm up holder with vac-lock) were used to each target. After deciding on the target¡¯s most stable respiratory cycles, the targets were asked to wear a goggle monitor and maintain their respiration regularly for thirty minutes to obtain the respiratory signals. To analyze the respiratory signal, the standard deviation and the variation value of the peak value and the valley value of the respiratory signal were separated by time zone with the self-developed program at the hospital and each tie-downs were compared for the estimation by calculating a comparative index using the above.

Results: The stability of each immobilizations were measured in consideration of deviation changes studied in each respiratory time lapse. Comparative indexes of each immobilizations of each experimenter are shown to be A: 11.20, B: 4.87, C: 1.63 / A: 3.94, B: 0.67, C: 0.13 / A: 2.41, B: 0.29, C: 0.04 / A: 0.16, B: 0.19, C: 0.007 / A: 35.70, B: 2.37, C: 1.86. And when all five experimenters wore the immobilizations C, the test proved the most stable value while four people wearing A and one man wearing D expressed relatively the most unstable respiratory outcomes.

Conclusion: The self-developed immobilizations, so called the arm up holder vac-lock for the stereotactic body radiotherapy is expected to improve the effect of the treatment by decreasing the intra-fraction organ motions because it keeps the respiration more stable than other two immobilizations. Particularly in case of the stereotactic body therapy which requires the maintenance of set-up state for a long time, the self-developed immobilizations is thought to more useful for stereotactic body radiotherapy rather than the rest two immobilizations with instable respiratory cycle as time passes.
KEYWORD
stereotactic body radiation therapy, immobilizations, respiratory signal