KMID : 1164220110230020097
|
|
Journal of Korean Society for Radiotherapeutic Technology 2011 Volume.23 No. 2 p.97 ~ p.102
|
|
Evaluation of Uncertainty of IMRT QA Using 2Dimensional Array Detector for Head & Neck Patients
|
|
Ban Tae-Joon
Lee Woo-Suk Kim Dae-Sup Back Geum-Mun Kwak Jung-Won
|
|
Abstract
|
|
|
Purpose: IMRT QA using 2Dimensional array detector is carried out with condition for discrete dose distribution clinically. And it can affect uncertainty of evaluation using gamma method. We analyze gamma index variation according to grid size and suggest validate range of grid size for IMRT QA in Hospital.
Materials and Methods: We performed QA using OniPro I¡¯mRT system software version 1.7b on 10 patients (head and neck) for IMRT. The reference dose plane (grid size, 0.1 cm; location, [0, 0, 0]) from RTP was compared with the dose plane that has different grid size (0.1 cm, 0.5 cm, 1.0 cm, 2.0 cm, 4.0 cm) and different location (along Y-axis 0 cm, 0.2 cm, 0.5 cm, 1.0 cm). The gamma index variation was evaluated by observing the level of changes in Gamma pass rate, Average signal, Standard deviation for each case.
Results: The average signal for each grid size showed difference levels of 0%, -0.19%, -0.04%, -0.46%, -8.32% and the standard deviation for each grid size showed difference levels of 0%, -0.30%, 1.24%, -0.70%, -7.99%. The gamma pass rate for each grid size showed difference levels of 0%, 0.27%, -1.43%, 5.32%, 5.60%. The gamma evaluation results according to distance in grid size range of 0.1 cm to 1.0 cm showed good agreement with reference condition (grid size 0.1 cm) within 1.5% and over 5% in case of the grid size was greater than 2.0 cm.
Conclusion: We recognize that the grid size of gamma evaluation can make errors of IMRT QA. So we have to consider uncertainty of gamma evaluation according to the grid size and apply smaller than 2 cm grid size to reduce error and increase accuracy clinically.
|
|
KEYWORD
|
|
IMRT QA, gamma index, 2Dimensional array detector
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|