Examinations on Applications of Manual Calculation Programs on Lung Cancer Radiation Therapy Using Analytical Anisotropic Algorithm
Kim Jong-Min

Kim Dae-Sup
Hong Dong-Ki
Back Geum-Mun
Kwak Jung-Won
Abstract
Purpose: There was a problem with using MU verification programs for the reasons that there were errors of MU when using MU verification programs based on Pencil Beam Convolution (PBC) Algorithm with radiation treatment plans around lung using Analytical Anisotropic Algorithm (AAA). On this study, we studied the methods that can verify the calculated treatment plans using AAA.

Materials and Methods: Using Eclipse treatment planning system (Version 8.9, Varian, USA), for each 57 fields of 7 cases of Lung Stereotactic Body Radiation Therapy (SBRT), we have calculated using PBC and AAA with dose calculation algorithm. By developing MU of established plans, we compared and analyzed with MU of manual calculation programs. We have analyzed relationship between errors and 4 variables such as field size, lung path distance of radiation, Tumor path distance of radiation, effective depth that can affect on errors created from PBC algorithm and AAA using commonly used programs.

Results: Errors of PBC algorithm have showned 0.2¡¾1.0% and errors of AAA have showned 3.5¡¾2.8%. Moreover, as a result of analyzing 4 variables that can affect on errors, relationship in errors between lung path distance and MU, connection coefficient 0.648 (P=0.000) has been increased and we could calculate MU correction factor that is A.E=L.P 0.00903+0.02048 and as a result of replying for manual calculation program, errors of 3.5¡¾2.8% before the application has been decreased within 0.4¡¾2.0%.

Conclusion: On this study, we have learned that errors from manual calculation program have been increased as lung path distance of radiation increases and we could verified MU of AAA with a simple method that is called MU correction factor.
KEYWORD
PBC, AAA, manual calculation, lung cancer