KMID : 1164220140260010077
Journal of Korean Society for Radiotherapeutic Technology
2014 Volume.26 No. 1 p.77 ~ p.81
The evaluation of lens absorbed dose according to the Optimold for whole brain radiation therapy


Park Byoung-Suk
Ahn Jong-Ho
Song Gi-Won
Abstract
Purpose : In the current whole brain Radiation Therapy, Optimold was used to immobilize the head. However, skin dose was increased about 22% due to the scattering radiation by the Optimold. Since the minimum dose causing cataracts was 2 Gy, it could be seen that the effects were large especially on the lens. Therefore, in the whole brain Radiation Therapy, it was to compare and to evaluate the lens absorbed dose according to the presence of Optimold in the eyeball part.

Materials and Methods : In order to compare and to evaluate the lens absorbed dose according to the presence of Optimold in the eyeball part, the Optimold mask was made ??up to 5mm bolus on the part of the eye lens in the human model phantom (Anderson Rando Phantom, USA). In the practice treatment, to measure the lens dose, the simulation therapy was processed by placing the GafChromic EBT3 film under bolus, and after the treatment plan was set up through the treatment planning system (Pinnacle, PHILIPS, USA), the treatments were measured repeatedly three times in the same way. After removing the Optimold mask in the eyeball part, it was measured in the same way as above. After scanning the film and measuring the dose by using the Digital Flatbed Scanner (Expression 10000XL, EPSON, USA), the doses were compared and evaluated according to the presence of Optimold mask in the eyeball part.

Results : When there was the Optimold mask in the eyeball part, it was measured at 10.2 cGy ¡¾ 1.5 in the simulation therapy, and at 24.8 cGy ¡¾ 2.7 in the treatment, and when the Optimold mask was removed in the eye part, it was measured at 12.9cGy ¡¾ 2.2 in the simulation therapy, and at 17.6 cGy ¡¾ 1.5 in the treatment.

Conclusion : In case of removing the Optimold mask in the eyeball part, the dose was increased approximately 3 cGy in the simulation therapy and was reduced approximately 7 cGy in the treatment in comparison to the case that the Optimold mask was not removed. During the whole treatment, since the lens absorbed dose was reduced about 27%, the chance to cause cataracts and side effects was considered to be reduced due to decrease of the absorbed dose to the eye lens which had the high sensitivity on the radiation.
KEYWORD
Whole brain radiation therapy, lens, GAFCHROMIC EBT3 film, Optimold