KMID : 1164220140260020329
Journal of Korean Society for Radiotherapeutic Technology
2014 Volume.26 No. 2 p.329 ~ p.336
The evaluation of contralateral breast¡¯s dose and shielding efficiency by breast size about breast implant patient for radiation therapy
Kim Jong-Wook

Woo Heon
Jeong Hyeon-Hak
Kim Kyeong-Ah
Kim Chan-Yong
Yoo Suk-Hyun
Abstract
Purpose : To evaluate the dose on a contralateral breast and the usefulness of shielding according to the distance between the contralateral breast and the side of the beam by breast size when patients who got breast implant receive radiation therapy.

Materials and Methods : We equipped 200 cc, 300 cc, 400 cc, and 500 cc breast model on the human phantom (Rando-phantom), acquired CT images (philips 16channel, Netherlands) and established the radiation treatment plan, 180 cGy per day on the left breast (EclipseTM ver10.0.42, Varian Medical Systems, USA) by size. We set up each points, A, B, C, and D on the right(contralateral) breast model for measurement by size and by the distance from the beam and attached MOSFET at each points. The 6 MV, 10 MV and 15 MV X-ray were irradiated to the left(target) breast model and we measured exposure dose of contralateral breast model using MOSFET. Also, at the same condition, we acquired the dose value after shielding using only Pb 2 mm and bolus 3 mm under the Pb 2 mm together.

Results : As the breast model is bigger from 200 cc to 500 cc, The surface of the contralateral breast is closer to the beam. As a result, from 200 cc to 500 cc, on 180 cGy basis, the measurement value of the scattered ray inclined by 3.22 ~ 4.17% at A point, 4.06 ~ 6.22% at B point, 0.4~0.5% at C point, and was under 0.4% at D point. As the X-ray energy is higher, from 6 MV to 15 MV, on 180 cGy basis, the measurement value of the scattered ray inclined by 4.06~5% at A point, 2.85~4.94% at B point, 0.74~1.65% at C point, and was under 0.4% at D point. As using Pb 2 mm for shield, scattered ray declined by average 9.74% at A and B point, 2.8% at C point, and is under 1% at D point. As using Pb 2 mm and bolus together for shield, scattered ray declined by average 9.76% at A and B point, 2.2% at C point, and is under 1% at D point.

Conclusion : Commonly, in case of patients who got breast implant, there is a distance difference by breast size between the contralateral breast and the side of beam. As the distance is closer to the beam, the scattered ray inclined. At the same size of the breast, as the X-ray energy is higher, the exposure dose by scattered ray tends to incline. As a result, as low as possible energy wihtin the plan dose is good for reducing the exposure dose.