KMID : 1164220150270010073
Journal of Korean Society for Radiotherapeutic Technology
2015 Volume.27 No. 1 p.73 ~ p.78
Evaluation of difference in respiratory phase between amplitude- and phase-based four-dimensional computed tomography
Lee So-Hyang

Park Su-Yeon
Kim Jong-Sik
Choi Byeong-Ki
Park Hee-Chul
Jung Sang-Hoon
Abstract
Purpose : Under the assumption of change to the amplitude based sorting, the study will use four dimensional computed tomography imaging (4DCT) arrayed using the phase based sorting to analyze the respiratory phase difference.

Materials and Methods : The study analyzed the 4DCT (4-dimensional computed tomography) images of 10 liver cancer patients that were treated with respiratory gated radiotherapy from 2015 February to March. Using RPM respiratory gating (RPM 1.7.5, Varian, USA) equipment, imaging according to respiratory cycle of phase based sorting was acquired and using a treatment planning system (Pinnacle 9.2, Philips, USA) the acquired imaging according to respiratory cycle was used to measure the abdominal movement value by respiratory cycle. The measuring point was the point where the center point of the Marker Block and the body surface met in the 50% phase image and here the coordinate values Lateral, Vertical, Longitudinal (X, Y, Z) were set as reference points, and on the X, Z plane identical to the reference point, using the identical method the Y axis coordinate value of each 0%, 30%, 40%, 50%, 60%, 80% phase images were acquired to quantitatively measure the variation of distance to the Y axis. The abdominal movement value according to respiration was applied to the theoretical model that the value decreases linearly from maximum inhalation to maximum exhalation to divide the variation of my value to predict as amplitude value by respiratory cycle and conversely the variation in amplitude was recalculated with the phase variation deviation value to analyze.

Results : The deviation value between expected value and actual location was the largest in the 30% phase with 0.24 cm, and standard deviation was also the largest in 30% phase with 0.13 cm. The effective value of the deviation value derived from the average of the deviation squared value of each patient appeared as minimum 0.7 cm, maximum 0.18 cm, average 0.12 cm, and standard deviation 0.4 cm. Also by dividing the actual movement distance value with the peak expiration value then converting it into %Phase, the deviation value with actual phase 16.5% in 30% phase, 10.0% and 40% phase, 10.0% and 60% phase, 15.4% and 80% phase, and overall average about 13%, and arraying based on amplitude, phase shift occurred and further it was from peak expiration the chance of deviation occurrence was increasingly measured.

Conclusion : Based on the results of the study there were differences between value acquired based on theoretical model and actual value. Therefore in respiratory gated radiotherapy using external surrogates, there needs to be establishment of respiration gated radiation system that avoids the combination of two Sorting methods considering that there will be occurrence of treatment and corresponding clinical differences due to the phase difference that occur due to the Amplitude based Phase Sorting.
KEYWORD
4DCT, Respiratory Motion, Respiratory Gated Radiotherapy, Amplitude based Sorting, Phase Based Sorting