Journal of Korean Society for Radiotherapeutic Technology
2015 Volume.27 No. 2 p.115 ~ p.122
Evaluation of the accuracy of the HexaPOD evo RT system using Non-coplanar beams in lung cancer
Chang Se-Wuk

Cho Kang-Chul
Lee Sang-Kyu
Kim Joo-Ho
Cho Jeong-Hee
Abstract
Purpose : The aim of this study, evaluate the accuracy of HeaxPOD evo RT system using the non-coplanar beam.

Materials and Methods : 13 treatment plans are used which applied non-coplanar beams and 10 treatment plans which
coplanar beams are used. the correction value what adjust to 6D couch is determined by each patient¡¯s setup errors only
rotation direction. The study executed followings. first, Applying the correction value, measure the point dose and calculate the ¥ã-index(¥ã=3% / 3 mm , ¥ã=2%/ 2 mm). second, acquire data as previous methods without correction by HexaPOD.

Results : For comparing the two results, we find out the more precise applying HexaPOD by point dose 0.2% in coplanar and non-coplanar. in the case of ¥ã-index<1(¥ã=3% / 3 mm), more precise 2.2% in coplanar and 7% in Non-coplanar. Particularly, ¥ã-index<1(2% / 2 mm) show the difference 9.2% in coplanar and 15.1% non-coplanar between apply HexaPOD and dose not apply HexaPOD.

Conclusion : Using the HexaPOD is more precise than without HexaPOD. It suggests that HexaPOD evo RT system is very
useful for precise and high dose delivery.