Journal of Korean Society for Radiotherapeutic Technology
2015 Volume.27 No. 2 p.131 ~ p.143
A study of usefulness for the plan based on only MRI using ViewRay MRIdian system
Jeon Chang-Woo

Lee Ho-Jin
Ahn Bum-Suk
Kim Chan-Yong
Lee Je-Hee
Abstract
Purpose : By comparing a CT fusion plan based on MRI with a plan based on only MRI without CT, we intended to study usefulness of a plan based on only MRI. And furthermore, we intended to realize a realtime MR-IGRT by MRI image without CT scan during the course of simulation, treatment planning, and radiation treatment.

Materials and Methods : BBB CT (Brilliance Big Bore CT, 16slice, Philips), Viewray MRIdian system (Viewray, USA) were used for CT & MR simulation and Treatment plan of 11 patients (1 Head and Neck, 5 Breast, 1 Lung, 3 Liver, 1 Prostate). When scanning for treatment, Free Breathing was enacted for Head&Neck, Breast, Prostate and Inhalation Breathing Holding for Lung and Liver. Considering the difference of size between CT and Viewray, the patient's position and devices were in the same condition. Using Viewray MRIdian system, two treatment plans were established. The one was CT fusion treatment plan based on MR image. Another was MR treatment plan including electron density that [ICRU 46] recommend for Lung, Air and Bone. For Head&Neck, Breast and Prostate, IMRT was established and for Lung and Liver, Gating treatment plan was established. PTV's Homogeneity Index(HI) and Conformity Index(CI) were use to estimate the treatment plan. And DVH and dose difference of each PTV and OAR were compared to estimate the treatment plan.

Results : Between the two treatment plan, each difference of PTV's HI value is 0.089% (Head&Neck), 0.26% (Breast), 0.67% (Lung), 0.2% (Liver), 0.4% (Prostate) and in case of CI, 0.043% (Head&Neck), 0.84% (Breast), 0.68% (Lung), 0.46% (Liver), 0.3% (Prostate). As showed above, it is on Head&Neck that HI and CI's difference value is smallest. Each difference of average dose on PTV is 0.07 Gy (Head&Neck), 0.29 Gy (Breast), 0.18 Gy (Lung), 0.3 Gy (Liver), 0.18 Gy (Prostate). And by percentage, it is 0.06% (Head&Neck), 0.7% (Breast), 0.29% (Lung), 0.69% (Liver), 0.44% (Prostate). Likewise, All is under 1%. In Head&Neck, average dose difference of each OAR is 0.01~0.12 Gy, 0.04~0.06 Gy in Breast, 0.01~0.21 Gy in Lung, 0.06~0.27 Gy in Liver and 0.02~0.23 Gy in Prostate.

Conclusion : PTV's HI, CI dose difference on the Treatment plan using MR image is under 1% and OAR's dose difference is maximum 0.89 Gy as heterogeneous tissue increases when comparing with that fused CT image. Besides, It characterizes excellent contrast in soft tissue. So, radiation therapy using only MR image without CT scan is useful in the part like Head&Neck, partial breast and prostate cancer which has a little difference of heterogeneity.
KEYWORD
MRI only based plan, Real time MR-IGRT