Journal of Korean Society for Radiotherapeutic Technology
2015 Volume.27 No. 2 p.175 ~ p.181
Contralateral Breast Doses Depending on Treatment Set-up Positions for Left-sided Breast Tangential Irradiation
Joo Chan-Seong

Park Su-Yeon
Kim Jong-Sik
Choi Byeong-Ki
Chung Yoon-Sun
Park Won
Abstract
Purpose : To evaluate Contralateral Breast Doses with Supine and Prone Positions for tangential Irradiation techniques for left-sided breast Cancer.

Materials and Methods : We performed measurements for contralateral doses using Human Phantom at each other three plans (conventional technique, Field-in-Field, IMRT, with prescription of 50 Gy/25fx). For the measurement of contralateral doses we used Glass dosimeters on the 4 points of Human Phantom surface (0 mm, 10 mm, 30 mm, 50 mm). For the position check at every measurements, we had taken portal images using EPID and denoted the incident points on the human phantom for checking the constancy of incident points.

Results : The contralateral doses in supine position showed a little higher doses than those in prone position. In the planning study, contralateral doses in the prone position increased mean doses of 1.2% to 1.8% at each positions while those in the supine positions showed mean dose decreases of 0.8% to 0.9%. The measurements using glass dosimeters resulted in dose increases (mean: 2.7%, maximum: 4% of the prescribed dose) in the prone position. In addition, the delivery techniques of Field-in-field and IMRT showed mean doses of 3% higher than conventional technique.

Conclusion : We evaluated contralateral breast doses depending on different positions of supine and prone for tangential irradiations. For the phantom simulation of set-up variation effects on contralateral dose evaluation, although we used humanoid phantom for planning and measurements comparisons, it would be more or less worse set-up constancy in a real patient. Therefore, more careful selection of determination of patient set-up for the breast tangential irradiation, especially in the left-sided breast, should be considered for unwanted dose increases to left lung and heart. In conclusion, intensive patient monitoring and improved patient set-up verification efforts should be necessary for the application of prone position for tangential irradiation of left-sided breast cancer.
KEYWORD
left-sided breast, tangential irradiation, prone position