Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1164220160280010047
Journal of Korean Society for Radiotherapeutic Technology
2016 Volume.28 No. 1 p.47 ~ p.55
Analysis of dosimetric leaf gap variation on dose rate variation for dynamic IMRT
Yang Myung-Sic

Park Ju-Kyeong
Lee Seung-Hun
Kim Yang-Su
Lee Sun-Young
Cha Seok-Yong
Abstract
To evaluate the position accuracy of the MLC. This study analyzed the variations of the dosimetric leaf gap(DLG) and MLC transmission factor to reflect the location of the MLC leaves according to the dose rate variation for dynamic IMRT.
We used the 6 MV and 10 MV X-ray beams from linear accelerator with a Millennium 120 MLC system. We measured the
variation of DLG and MLC transmission factor at depth of 10 cm for the water phantom by varying the dose rate to 200, 300,
400, 500 and 600 MU/min using the CC13 and FC-65G chambers. For 6 MV X-ray beam, a result of measuring based on a dose rate 400 MU/min by varying the dose rate to 200, 300, 400, 500 and 600 MU/min of the difference rate was respectively -2.59, -1.89, 0.00, -0.58, -2.89%. For 10 MV X-ray beam, the difference rate was respectively ?2.52, -1.69, 0.00, +1.28, -1.98%. The difference rate of MLC transmission factor was in the range of about ¡¾1% of the measured values at the two types of energy and all of the dose rates. This study evaluated the variation of DLG and MLC transmission factor for the dose rate variation for dynamic IMRT. The difference of the MLC transmission factor according to the dose rate variation is negligible, but, the difference of the DLG was found to be large. Therefore, when randomly changing the dose rate dynamic IMRT, it may significantly affect the dose delivered to the tumor. Unless you change the dose rate during dynamic IMRT, it is thought that is to be the more accurate radiation therapy.
KEYWORD
IMRT, Dosimetric leaf gap, MLC transmission factor, Dose rate
FullTexts / Linksout information
 
Listed journal information