KMID : 1164220160280020109
|
|
Journal of Korean Society for Radiotherapeutic Technology 2016 Volume.28 No. 2 p.109 ~ p.121
|
|
Evaluation of accuracy in the ExacTrac 6D image induced radiotherapy using CBCT
|
|
Park Ho-Choon
Kim Hyo-Jung Kim Jong-Deok Ji Dong-Hwa Song Ju-Young
|
|
Abstract
|
|
|
To verify the accuracy of the image guided radiotherapy using ExacTrac 6D couch, the error values in six directions are randomly assigned and corrected and then the corrected values were compared with CBCT image to check the accurateness of ExacTrac. The therapy coordination values in the Rando head Phantom were moved in the directions of X, Y and Z as the translation group and they were moved in the directions of pitch, roll and yaw as the rotation group. The corrected values were moved in 6 directions with the combined and mutual reactions. The Z corrected value ranges from 1mm to 23mm. In the analysis of errors between CBCT image of the phantom which is corrected with therapy coordinate and 3D/3D matching error value, the rotation group showed higher error value than the translation group. In the distribution of dose for the error value of the therapy coordinate corrected with CBCT, the restricted value of dosage for the normal organs in two groups meet the prescription dose. In terms of PHI and PCI values which are the dose homogeneity of the cancerous tissue, the rotation group showed a little higher in the low dose distribution range. This study is designed to verify the accuracy of ExacTrac 6D couch using CBCT. It showed that in terms of the error value in the simple movement, it showed the comparatively accurate correction capability but in the movement when the angle is put in the couch, it showed the inaccurate correction values. So, if the body of the patient is likely to have a lot of changes in the direction of rotation or there is a lot of errors in the pitch, roll and yaw in ExacTrac correction, it is better to conduct the CBCT guided image to correct the therapy coordinate in order to minimize any side effects.
|
|
KEYWORD
|
|
ExacTrac 6D couch , image guided radiotherapy , CBCT , pitch , roll and yaw
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|