Journal of Korean Society for Radiotherapeutic Technology
2017 Volume.29 No. 1 p.57 ~ p.68
Comparison of Volumetric Modulated Arc Therapy and Non-coplanar Fixed-field Intensity Modulated Radiation Therapy for Irregular Target adjacent to Organ At Risk
Kim Kyeong-Ah

Na Kyoung-Su
Seo Seok-Jin
Lee Je-Hee
Abstract
Purpose: The purpose of this study was tocompare volumetric modulated arc therapy(VMAT) with fixed-field intensity modulated radiation therapy(IMRT) using non-coplanar beam when the shape of target is irregular and the location is adjacent to organ at risk(OAR).

Materials and Methods: The subjects of this study were a total of 6 patients who had radiation therapy for whole scalp(2 patients), partial scalp(2 patients), and whole ventricle(2 patients) by True Beam STX(Varian Medical Systems, USA). VMAT plans consisted of coplanar or non-coplanar arcs which can minimize the volume of OAR included in beamlets. All fixed-field IMRT plans consisted of non-coplanar beams using more than 2 angles of Couch.

Results: The VMAT and IMRT plans were compared with regard to the maximum dose of both lens, both optic nerves, optic chiasm, and brain stem and the mean dose of both eyeballs and hippocampus. VMAT plans showed higher dose than ncIMRT plans at more than 6 of all OARs in every patient, and the ratio was from 1.1 times to 8.2 times. In case of total scalp and partial scalp, the volume of brain which received more than 20 Gy in the VMAT plans was 2 times larger than the volume in the ncIMRT plans. In case of whole ventricle, there was no significant difference. Target coverage was satisfied in both plans PTV 100%=95%. The maximum dose in target volume and required monitor unit(MU) of ncIMRT were higher than them of VMAT plans.

Conclusion: Even though ncIMRT is less efficient than VMAT with regard to required MU and treatment time, the dose to OARs is much lower than VMAT and PTV Coverage is similar with VMAT. If the shape of target is irregular and location is adjacent to OAR, comparison VMAT plan with ncIMRT plan deserves to be considered.
KEYWORD
Irregular shape of target, Organ At Risk, Non-coplanar Beam, Intensity Modulated Radiation Therapy, Volumetric Modulated Arc Therapy, Whole Scalp, Partial Scalp, Whole Ventricle