Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1164220170290010069
Journal of Korean Society for Radiotherapeutic Technology
2017 Volume.29 No. 1 p.69 ~ p.76
A study on dosimetric comparison of craniospinal irradiation using tomotherpy and reproducibility of position
Lee Hee-Jeong

Kim Joo-Ho
Lee Sang-Kyu
Yoon Jong-Won
Cho Jeong-Hee
Abstract
Purpose: The purpose of this study was to verify dosimetric results and reproducibility of position during craniospinal irradiation (CSI) using tomotherapy (Accuray Incorporated, USA). Also, by comparing with conventional CSI Technique, we confirmed the efficiency of using a Tomotherapy.

Materials and Methods: 10 CSI patients who get tomotherapy participate. Patient-specific quality assurances (QA) for each patient are conducted before treatment. When treating, we took Megavoltage Computed Tomography (MVCT) that range of head and neck before treatment, L spine area after treatment. Also we conducted in-vivo dosimetry to check a scalp dose. Finally, we made a 3D conventional radiation therapy(3D-CRT) of those patients to compare dosimetric differences with tomotherapy treatment planning.

Results: V107, V95 of brain is 0 %, 97.2 % in tomotherapy, and 0.3 %, 95.1 % in 3D-CRT. In spine, value of V107, V95 is 0.2 %, 18.6 % in tomotherapy and 89.6 %, 69.9 % in 3D-CRT. Except kidney and lung, tomotherapy reduced normal organ doses than 3D-CRT. The maximum positioning error value of X, Y, Z was 10.2 mm, -8.9 mm, -11.9 mm. Through in-vivo dosimetry, the average of scalp dose was 67.8 % of prescription dose. All patient-specific QA were passed by tolerance value.

Conclusion: CSI using tomotherapy had a risk of parallel organ such as lung and kidney because of integral dose in low dose area. However, it demonstrated dosimetric superiority at a target and saved normal organ to reduce high dose. Also results of reproducibility were not exceeded margins that estimated treatment planning and invivo dosimetry showed to reduce scalp dose. Therefore, CSI using tomotherapy is considered to efficient method to make up for 3D-CRT.
KEYWORD
Craniospinal Irradiation, Tomotherapy
FullTexts / Linksout information
 
Listed journal information